Olive Non-invasive Positive Pressure Ventilator is designed for precision, comfort, and ease of use, making it ideal for both medical institutions and home care settings. Whether for hospitals, clinics, or home healthcare, our ventilator delivers professional-grade therapy with user-friendly operation, ensuring better patient outcomes and peace of mind.
What is Non-Invasive Positive Pressure Ventilator?
A Non-Invasive Positive Pressure Ventilator (NIPPV) is a respiratory support device that delivers pressurized air through a mask, assisting patients with conditions such as COPD, respiratory insufficiency, and sleep apnea. Unlike invasive ventilators, which require intubation, NIPPV offers a safer, more comfortable alternative by reducing infection risks and enhancing patient compliance. It ensures adequate ventilation by maintaining airway pressure, improving oxygenation, and reducing the work of breathing.
Dimension | 226x196x175 mm |
Weight | 2.0 kg without battery 2.4 kg with battery |
Environment temperature | Working environment: +5°C to+40°C Transportation and storage: -20°C to+55°C |
Humidity range | Working environment: 10%~95% Transportation and storage: ≤ 93% |
Atmospheric pressure range | 60kPa-110kPa |
Anti-electric shock type | Class ll device |
Degree ofAnti-electric shock | BF type application part |
Degree of protection against harmful liquid inlet | IPX1 |
Operating noise | ≤30dB (CPAP 10 + 4mm standard plug) |
Operation mode | continuous operation |
Normal working rated power | 80W |
External power supply | AC 100-240V,50-60HZ,2A |
Built-in battery specifications | 18V,6000mAh,108Wh |
Built-in battery model | HK-sam30g-5s2p15a |
Running time * | 8 hours when the built-in battery is fully charged *CPAP mode, 10cmH20 pressure, running time is greater than or equal to 8 hours |
Work Mode* | CPAP, S,T, ST, APCV,TVV-ST, TVV-APCV, IPVS, HF *IPVS and HF modes are model P device modes |
Inspiratory pressure (IPAP) | 4cmH₂O~40cmH₂O 4-40cmH₂O 4cmH₂O~35cmH₂O 4-35cmH₂O 4cmH₂O~30cmH₂O 4-30cmH₂O * Inspiratory pressure Settings for S, T, ST, APCV, TVVST, and TVVAPCV modes.The setting value of the maximum inspiratory pressure is different for different models |
Expiratory pressure(EPAP) | 4cmH₂O~40cmH₂O 4cmH₂O~35cmH₂O 4cmH₂O~30cmH₂O * Expiratory pressure setting value of S, T, ST, APCV, TVV-ST, TVV-APCV modesThe setting value of maximum expiratory pressure is different for different models |
Pressure adjustment interval | 0.2cmH₂O |
Pressure accuracy | ±0.5cmH₂O or 4% set value; Take the large value of both |
Maximum inspiratory pressure * (Max IPAP) | 4cmH₂O~40cmH₂O 4cmH₂O~35cmH₂O 4cmH₂O~30cmH₂O * Maximum inspiratory pressure Settings for APCV, TVV-ST, and TVV-APCV modes |
Minimum inspiratory pressure *( Min IPAP ) | 4cmH₂O~40cmH₂O 4cmH₂O~35cmH₂O 4cmH₂O~30cmH₂O * Minimum inspiratory pressure Settings for APCV, TVV-ST, and TVV-APCV modes,which are ess than or eaua to the maximum inspiratory pressure |
Pressure rise/drop duration | 100ms-900ms |
Backup Rate | 5-60 times/min(T mode) |
I/E Rate | 10%-90%(T mode) |
IPAP/Inspiratory Sensitivity | Very high, high, medium, low, very low |
Target tidal volume | 200-1200ml |
Inspiratory time | 0.1-4s |
Min/Max inspiratory time | 0.1-4s |
Ramp time | 0-60min 1min Step length |
Max flow rate | 5cmH₂O-190L/min 10cmH₂O-180L/min 15cmH₂O-170L/min 20cmH₂O-165L/min 35cmH₂O-130L/min(RX-35 and RX-40 models only) 40cmH₂O-30L/min(RX-40 model only) |
HF high flow mode | Output flow setting range: 10-80L/min Step:1 L/min Accuracy: +10% or 2L/min; Take the large value of both Backup respiratory rate: 5bpm-60bpm Step 1 BPM |
Max ultimate pressure | RX-30: max ultimate pressure ≤35cmH₂O (≈3.5kPa) in a single fault state, RX-35: max ultimate pressure ≤40cmH₂O (≈4.0kPa) in a single fault state. RX-40: max ultimate pressure ≤45cmH₂O (≈4.5kPa) in a single fault state; If Max ultimate pressure is exceeded for more than 10 seconds, the device will automatically turn off the blower and stop the pressure output. |